Abstrakt
The paper presents a numerical method of blood flow simulation in two geometries. The first geometry covered the ascending and descending parts of aorta and its main branches in patient with a congenital heart disease-coarctation of aorta (inborn narrowing of the descendent aorta). The second geometry used in the flow field analysis mimicked the state of the vessel after a successful percutaneous intervention and thus without pathological narrowing (virtual therapy). The blood was treated as a non-Newtonian fluid. The pulsating flow was set at the inlet to ascending aorta. The lumped model based on electrical analogy using three-element-Windkessel model was created to calculate pressure at the outlets. In the publication the fields of velocity and pressure were presented. The obtained outflow streams were modified after virtual therapy.
| Język oryginału | angielski |
|---|---|
| Strony (od–do) | 153-160 |
| Liczba stron | 8 |
| Czasopismo | Advances in Intelligent Systems and Computing |
| Tom | 526 |
| Identyfikatory DOI | |
| Status publikacji | Opublikowano - 1 paź 2017 |