Abstrakt
The main thesis of the research was the problem concerning calibration of high-temperature solution calorimeter with aluminum. The important assumptions applied at calibration of a device with aluminum were verified. Enthalpy values determined during the experiment clearly suggest that the level of deformation and diameters of aluminum samples used for calibration are negligible. The assumption that the temperature of the samples at the moment they reach the bath level is very close to room temperature, is appropriate.