The studies of assumptions accompanying the calibration of high-temperature solution calorimeter

  • Polish Academy of Sciences
  • Silesian University of Technology

Wyniki badań: Wkład do czasopisma › Artykuł › recenzja

5 Cytowania z bazy Scopus

Abstrakt

The main thesis of the research was the problem concerning calibration of high-temperature solution calorimeter with aluminum. The important assumptions applied at calibration of a device with aluminum were verified. Enthalpy values determined during the experiment clearly suggest that the level of deformation and diameters of aluminum samples used for calibration are negligible. The assumption that the temperature of the samples at the moment they reach the bath level is very close to room temperature, is appropriate.

Język oryginałuangielski
Strony (od–do)585-592
Liczba stron8
CzasopismoArchives of Metallurgy and Materials
Tom56
Numer wydania3
Identyfikatory DOI
Status publikacjiOpublikowano - 2011