Abstrakt
This paper addresses issues relating to the temperature stability of impedance infrastructure used for the realization of impedance units and the calibration of measuring instruments. A description of the new infrastructure for impedance metrology recently developed in Poland is provided. We present the results of measurements of temperature coefficients for four-port resistance standards and two-port capacitors used in the permuting capacitor device. We also report the results of stability tests of the recently developed temperature-controlled capacitance set. The results demonstrate excellent temperature stability for the impedance components and serve as an important source of information for selecting environmental conditions during comparisons and routine calibration.
| Język oryginału | angielski |
|---|---|
| Numer artykułu | 3018 |
| Czasopismo | Energies |
| Tom | 18 |
| Numer wydania | 12 |
| Identyfikatory DOI | |
| Status publikacji | Opublikowano - cze 2025 |
Obszary tematyczne ASJC Scopus
- Energia odnawialna, zrównoważony rozwój i środowisko
- Technologia paliwowa
- Inżynieria (różne)
- Inżynieria energetyczna i technologia energetyczna
- Energia (różne)
- Sterowanie i optymalizacja
- Inżynieria elektryczna i elektroniczna
Fingerprint
Zanurz się w tematy badawcze publikacji „Temperature Dependence of the New Calibration Infrastructure for Impedance Metrology”. Razem tworzą niepowtarzalny odcisk palca.Cytowanie