N2 - ALD is a variation of CVD, which, for example, found application in depositing two-component compounds, such as SiO2 or TiO2. In the case of layers applied to the surfaces of products intended for contact with blood, important determinants include, aside from chemical composition, achieving an appropriate thickness and adequate sealing.Asecure layer of the right thickness forms an effective barrier that protects the metal nanomaterial from the effects of corrosive environments (Shan et al. in Surf Coat Technol 202:2399–2402, 2008[1]). Thin oxide layers based on such elements such as Ti or Si (up to 250 nm) are more hemocompatible, which significantly reduces the risk of complications related to the disseminated intravascular coagulation (DIC) process, for example. Aside from improved hemocompatibility, another important issue related to creating surface layers is the ability to achieve an appropriate set of electrochemical parameters. For this reason, tests were performed concerning the electrochemical properties of the TiO2 layers deposited on the surfaces of AISI 316LVMsteel samples using ALD under varied process parameters. First, potentiodynamic and potentiostatic measurements were taken, which enabled pitting and crevice corrosion resistance to be assessed. Secondly, impedance measurements were performed to enable interpretation of the processes and phenomena occurring at the TiO2 layer—electrolyte (synthetic plasma) interface. Complementary examination of surface topography was performed using a scanning electron microscope (SEM).
AB - ALD is a variation of CVD, which, for example, found application in depositing two-component compounds, such as SiO2 or TiO2. In the case of layers applied to the surfaces of products intended for contact with blood, important determinants include, aside from chemical composition, achieving an appropriate thickness and adequate sealing.Asecure layer of the right thickness forms an effective barrier that protects the metal nanomaterial from the effects of corrosive environments (Shan et al. in Surf Coat Technol 202:2399–2402, 2008[1]). Thin oxide layers based on such elements such as Ti or Si (up to 250 nm) are more hemocompatible, which significantly reduces the risk of complications related to the disseminated intravascular coagulation (DIC) process, for example. Aside from improved hemocompatibility, another important issue related to creating surface layers is the ability to achieve an appropriate set of electrochemical parameters. For this reason, tests were performed concerning the electrochemical properties of the TiO2 layers deposited on the surfaces of AISI 316LVMsteel samples using ALD under varied process parameters. First, potentiodynamic and potentiostatic measurements were taken, which enabled pitting and crevice corrosion resistance to be assessed. Secondly, impedance measurements were performed to enable interpretation of the processes and phenomena occurring at the TiO2 layer—electrolyte (synthetic plasma) interface. Complementary examination of surface topography was performed using a scanning electron microscope (SEM).
KW - AISI 316LVM
KW - ALD
KW - EIS
KW - Potentiodynamic test
KW - SEM
KW - TiO
UR - https://www.scopus.com/pages/publications/84982918604
U2 - 10.1007/978-981-10-1602-8_24
DO - 10.1007/978-981-10-1602-8_24
M3 - Conference contribution
AN - SCOPUS:84982918604
SN - 9789811016011
T3 - Advanced Structured Materials
SP - 297
EP - 308
BT - Properties and Characterization of Modern Materials
A2 - Öchsner, Andreas
A2 - Altenbach, Holm
PB - Springer Verlag
T2 - 9th International Conference on Advanced Computational Engineering and Experimenting, ACE-X 2015
Y2 - 29 June 2015 through 2 July 2015
ER -