Przeskocz do nawigacji głównej Przeskocz do wyszukiwania Przeskocz do głównej treści

Stress distribution around a TKR implant: Are lab results consistent with observational studies?

  • Medical University of Silesia in Katowice

Wyniki badań: Wkład do czasopismaArtykułrecenzja

8 Cytowania z bazy Scopus

Abstrakt

Malalignment of Total Knee Replacement prosthesis has been reported to limit the implant survival time. We hypothesized that this may be secondary to excessive stress occurring at the bone-implant interface, resulting from abnormal load transfer across the knee joint. In this study, we conducted Finite Element Analysis of a geometrical model of the knee joint after Total Knee Replacement with different axial alignments. The calculated stresses and displacements were significantly higher with varus knee malalignment than with the valgus one. The stresses are not high enough to pose a serious risk of a crack or a fracture but might be responsible for chronic pain reported by some patients. In cases where optimal implant positioning is not possible, slight valgus malalignment might produce better results than varus.

Język oryginałuangielski
Strony (od–do)21-26
Liczba stron6
CzasopismoActa of Bioengineering and Biomechanics
Tom10
Numer wydania4
Status publikacjiOpublikowano - 2008

Obszary tematyczne ASJC Scopus

  • Biofizyka
  • Bioinżynieria
  • Biomateriały
  • Inżynieria biomedyczna

Fingerprint

Zanurz się w tematy badawcze publikacji „Stress distribution around a TKR implant: Are lab results consistent with observational studies?”. Razem tworzą niepowtarzalny odcisk palca.

Cytuj to