Abstrakt
The article presents, compares and discusses the results of determining the parameters of the gas chromatograph calibration curve for the internal standard method. The purpose of the calibration was to determine the concentration of ethanol in liquid food samples. From a set of 168 experimental results, 12 outliers were selected. From the obtained data, test sets containing the outlier results were constructed. The parameters of the calibration curve were determined by 2 classical methods (LS, WLS) and 3 methods resistant to the occurrence of gross errors: using winsorization of residual errors, using Huber's sensitivity function and using the method of the least median square of residual errors.
results obtained by the 5 methods were compared, conclusions were drawn.
| Przetłumaczony tytuł wkładu | Comparison of classical and robust methods for estimating the parameters of the linear processing equation in the gas chromatograph calibration procedure |
|---|---|
| Język oryginału | polski |
| Strony (od–do) | 204-207 |
| Liczba stron | 4 |
| Czasopismo | Przeglad Elektrotechniczny |
| Tom | 2023 |
| Numer wydania | 10 |
| Identyfikatory DOI | |
| Status publikacji | Opublikowano - 2023 |
Słowa kluczowe
- FID
- Gas chromatography
- calibration curve
- gross errors
- outliers
- robust estimation methods
Obszary tematyczne ASJC Scopus