title = "Detection of low amplitude fetal QRS complexes",
abstract = "The significance of the most important operations performed by the noninvasive systems for fetal heart rate determination is investigated. The method of template subtraction for maternal ECG suppression is compared to the method based on independent component analysis. The QRS detector based on the classical slope-responsive preprocessing competes with the one that employs normalized matched filtering for QRS enhancement. A small database containing the four-channel abdominal ECG signals with the marked positions of the fetal QRS complexes is prepared to enable the investigations. The performed experiments show the factors that have the greatest impact on the results of the fetal QRS detection, and an effective approach to cope the problem is proposed.",
author = "Marian Kotas and Janusz Jezewski and Tomasz Kupka and Krzysztof Horoba",
year = "2008",
doi = "10.1109/iembs.2008.4650278",
language = "English",
isbn = "9781424418152",
series = "Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08 - {"}Personalized Healthcare through Technology{"}",
publisher = "IEEE Computer Society",
pages = "4764--4767",
booktitle = "Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08",
address = "United States",
note = "30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08 ; Conference date: 20-08-2008 Through 25-08-2008",
}