The paper presents results of a trilateral comparison of a travelling AC voltage standard, which was a fused-silica NIST planar multijunction thin-film thermal converter. The AC-DC voltage transfer difference of the standard was measured at 2 V and selected frequencies from 1 MHz to 30 MHz against primary thermal AC voltage standards at SUT, RISE and Trescal.

Original languageEnglish
Title of host publicationCPEM 2018 - Conference on Precision Electromagnetic Measurements
PublisherInstitute of Electrical and Electronics Engineers Inc.
ISBN (Electronic)9781538609736
DOIs
Publication statusPublished - 19 Oct 2018
Event2018 Conference on Precision Electromagnetic Measurements, CPEM 2018 - Paris, France
Duration: 8 Jul 201813 Jul 2018

Publication series

NameCPEM 2018 - Conference on Precision Electromagnetic Measurements

Conference

Conference2018 Conference on Precision Electromagnetic Measurements, CPEM 2018
Country/TerritoryFrance
CityParis
Period8/07/1813/07/18

Keywords

  • AC-DC transfer
  • AC-DC transfer difference
  • key comparison
  • measurement standards
  • thermal converter

ASJC Scopus subject areas

  • Electrical and Electronic Engineering
  • Instrumentation
  • Atomic and Molecular Physics, and Optics
  • Radiation

Fingerprint

Dive into the research topics of 'Trilateral Comparison of a Planar Thin-Film Thermal AC Voltage Standard in Frequency Range 1 MHz-30 MHz'. Together they form a unique fingerprint.

Cite this