In this article is shown the research process of electromagnetically driven blood micropump winding shape as well the configuration of magnetic circuit influence on its generated electromagnetic torque. Analysis of the winding shape is intended to increase the electromagnetic torque values generated by the micropump and to assure more smooth torque characteristic versus angular displacement of blood chamber. This analysis is based on a spatial field model of an electromagnetically driven pulse micropump and have been made using the finite element method.