Abstract
The paper attempts to estimate the impact of disturbances on the commercial medical device. As a source of disturbances the signals in accordance with current and proposed regulations and the signals produced by GSM system have been chosen. The signals were briefly characterized in time and frequency domain. The obtained results shows that obligatory recommendations could be insufficient in the evaluation of immunity of commercial medical devices on signals produced by radiocommunication systems.
| Translated title of the contribution | Ocena wpływu zaburzeń emitowanych przez urządzenia telefonii komórkowej gsm na aparaturę elektrokardiograficzną |
|---|---|
| Original language | English |
| Pages (from-to) | 151-154 |
| Number of pages | 4 |
| Journal | Przeglad Elektrotechniczny |
| Volume | 88 |
| Issue number | 12 B |
| Publication status | Published - 2012 |
Keywords
- EMC of medical devices
- EMC tests
- Electromagnetic compatibility (EMC)
- Radiated emission immunity
ASJC Scopus subject areas
- Electrical and Electronic Engineering
Fingerprint