Skip to main navigation Skip to search Skip to main content

Stress distribution around a TKR implant: Are lab results consistent with observational studies?

  • Medical University of Silesia in Katowice

Research output: Contribution to journalArticlepeer-review

8 Citations (Scopus)

Abstract

Malalignment of Total Knee Replacement prosthesis has been reported to limit the implant survival time. We hypothesized that this may be secondary to excessive stress occurring at the bone-implant interface, resulting from abnormal load transfer across the knee joint. In this study, we conducted Finite Element Analysis of a geometrical model of the knee joint after Total Knee Replacement with different axial alignments. The calculated stresses and displacements were significantly higher with varus knee malalignment than with the valgus one. The stresses are not high enough to pose a serious risk of a crack or a fracture but might be responsible for chronic pain reported by some patients. In cases where optimal implant positioning is not possible, slight valgus malalignment might produce better results than varus.

Original languageEnglish
Pages (from-to)21-26
Number of pages6
JournalActa of Bioengineering and Biomechanics
Volume10
Issue number4
Publication statusPublished - 2008

Keywords

  • Finite element analysis
  • Malalignment
  • Total knee replacement

ASJC Scopus subject areas

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Stress distribution around a TKR implant: Are lab results consistent with observational studies?'. Together they form a unique fingerprint.

Cite this