Skip to main navigation Skip to search Skip to main content

Statistics of count numbers from a photomultiplier tube and its implications for error estimation

  • University of Bern

Research output: Contribution to journalArticlepeer-review

27 Citations (Scopus)

Abstract

We present investigations of counting statistics for five different luminescence-measuring systems. The dark current background and the counts obtained for a PM tube under constant illumination at various light intensities were measured. Where no divider is present in the counting system, the standard deviation of the number of counts exceeds the value predicted by a Poisson distribution but by different amounts in the different systems. This has implications for analysis of the uncertainties in the estimated equivalent doses and other quantitative analyses of luminescence measurements. We suggest how to account for the observed overdispersion of count numbers in uncertainty estimation.

Original languageEnglish
Pages (from-to)746-751
Number of pages6
JournalRadiation Measurements
Volume47
Issue number9
DOIs
Publication statusPublished - Sept 2012

Keywords

  • Counting statistics of PMT
  • OSL
  • Overdispersion
  • TL

ASJC Scopus subject areas

  • Radiation
  • Instrumentation

Fingerprint

Dive into the research topics of 'Statistics of count numbers from a photomultiplier tube and its implications for error estimation'. Together they form a unique fingerprint.

Cite this