The influence of surface modification and steam sterilisation process on the corrosion characteristics of wires made of X10CrNi18-8 steel used for production of guide wires in endourological treatment was analysed in the study. Corrosion tests were made in artificial urine solution with application of electrochemical testing system VoltaLab PGP201. Resistance to pitting corrosion was evaluated on the ground of registered anodic polarisation curves by means of potentiodynamic method. The tests comprised wires with surface that was ground, electrochemically polished and chemically passivated. Steam sterilisation was performed in autoclave Basic Plus by Mocom. The tests proved that surface modification improved corrosion characteristics of the wire to a great extent. The best corrosion resistance in artificial urine was observed in wire whose surface was successively ground, polished and chemically passivated. It was also proved that chemical passivation process has a favourable impact on corrosion resistance improvement under sterilisation conditions.

Original languageEnglish
Title of host publicationTechnologies and Properties of Modern Utility Materials XXI
PublisherTrans Tech Publications Ltd.
Pages163-166
Number of pages4
ISBN (Print)9783037859452
DOIs
Publication statusPublished - 2014
Event21st Conference on Technologies and Properties of Modern Utility Materials, TPMUM 2013 - Katowice, Poland
Duration: 17 May 201317 May 2013

Publication series

NameSolid State Phenomena
Volume212
ISSN (Print)1012-0394

Conference

Conference21st Conference on Technologies and Properties of Modern Utility Materials, TPMUM 2013
Country/TerritoryPoland
CityKatowice
Period17/05/1317/05/13

Keywords

  • Endourology
  • Potentiodynamic tests
  • Steam sterilisation
  • X10CrNi18-8 steel

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • General Materials Science
  • Condensed Matter Physics

Fingerprint

Dive into the research topics of 'Resistance to electrochemical corrosion under conditions of sterilisation of guide wires used in endourological treatment'. Together they form a unique fingerprint.

Cite this