Abstract
We have proposed and deployed prototyped multispectral capturing device for the ophthalmic diagnosis. The main component of the device are slit lamp, liquid crystal tunable filter and high sensitive monochrome camera. The applied hardware allows to obtain 21 spectral channels in the visible light range. The device has been calibrated based on the captured images of the colorchecker colors with known RGB and spectral values.. The proposed device is focused on the diagnoses of cancer, PEX, galucoma and diabetic retinopathy,.
| Translated title of the contribution | Diagnostyka okulistyczna na bazie wielospektralnego obrazowania galki ocznej |
|---|---|
| Original language | English |
| Pages (from-to) | 165-168 |
| Number of pages | 4 |
| Journal | Przeglad Elektrotechniczny |
| Volume | 87 |
| Issue number | 12 B |
| Publication status | Published - 2011 |