Skip to main navigation Skip to search Skip to main content

Electrochemical characterisation of X10CrNi 18-8 steel in artificial plasma

  • Silesian University of Technology

Research output: Contribution to journal › Article › peer-review

Abstract

Quality of applied guide wires conditions the course of treatment and its success to a great extent. In order to simulate conditions that can be found in blood vascular system samples were exposed to artificial blood plasma (T = 37 ± 1 °C) for the time of 8 h. In order to obtain information regarding physical and chemical properties of modified surface of wire made of X10CrNi 18-8 steel, electrochemical impedance spectroscopy (EIS) test and tests of chemical composition of the surface layer X-ray photoelectron spectroscopy (XPS) were made. On the ground of performed EIS and XPS tests, favourable impact of steam sterilisation process on corrosion resistance of X10CrNi 18-8 steel was observed only for the case when chemical passivation was applied prior to sterilisation.

Original languageEnglish
Pages (from-to)63-66
Number of pages4
JournalMetalurgija
Volume54
Issue number1
Publication statusPublished - 1 Jan 2015

Keywords

  • EIS
  • Steam sterilization
  • Wires for invasive cardiology
  • X10CrNi 18-8 steel
  • XPS

ASJC Scopus subject areas

  • Condensed Matter Physics
  • Mechanics of Materials
  • Metals and Alloys
  • Materials Chemistry

Fingerprint

Dive into the research topics of 'Electrochemical characterisation of X10CrNi 18-8 steel in artificial plasma'. Together they form a unique fingerprint.

Cite this