Abstract
Quality of applied guide wires conditions the course of treatment and its success to a great extent. In order to simulate conditions that can be found in blood vascular system samples were exposed to artificial blood plasma (T = 37 ± 1 °C) for the time of 8 h. In order to obtain information regarding physical and chemical properties of modified surface of wire made of X10CrNi 18-8 steel, electrochemical impedance spectroscopy (EIS) test and tests of chemical composition of the surface layer X-ray photoelectron spectroscopy (XPS) were made. On the ground of performed EIS and XPS tests, favourable impact of steam sterilisation process on corrosion resistance of X10CrNi 18-8 steel was observed only for the case when chemical passivation was applied prior to sterilisation.
| Original language | English |
|---|---|
| Pages (from-to) | 63-66 |
| Number of pages | 4 |
| Journal | Metalurgija |
| Volume | 54 |
| Issue number | 1 |
| Publication status | Published - 1 Jan 2015 |
Keywords
- EIS
- Steam sterilization
- Wires for invasive cardiology
- X10CrNi 18-8 steel
- XPS
ASJC Scopus subject areas
- Condensed Matter Physics
- Mechanics of Materials
- Metals and Alloys
- Materials Chemistry
Fingerprint
Dive into the research topics of 'Electrochemical characterisation of X10CrNi 18-8 steel in artificial plasma'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver