Abstract

This paper presents methods of calibration of digital sources of ac voltage of very low frequency (0.001-100 Hz) and of effective value from 0.5 to 5 V. This work deals primarily with the calibration methods of the digital source of the sinusoidal signal [digitally synthesized source (DSS)] which is approximated with the staircase waveform. The construction of the DSS and the calibration system is presented. The calibration methods are divided into dynamic and static. Of the dynamic methods, the thermal method and integrating sampling method are briefly presented. Of the static methods, the step and the peak-to-peak methods of calibration are briefly presented. The potential applications, basic features, and accuracy of all four methods are discussed. This paper also presents a comparison of results obtained using the methods of calibration described in this paper.

Original languageEnglish
Article number6363601
Pages (from-to)1615-1620
Number of pages6
JournalIEEE Transactions on Instrumentation and Measurement
Volume62
Issue number6
DOIs
Publication statusPublished - 2013

Keywords

  • AC voltage standard
  • ac-dc transfer
  • digital signal synthesis
  • thermal converter

ASJC Scopus subject areas

  • Instrumentation
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Comparison of nonquantum methods for calibration of the digital source of very-low-frequency AC voltage'. Together they form a unique fingerprint.

Cite this