Abstract
A method of checking coherence of comparison results of primary DC voltage group standards is described in the paper. Usefulness of the proposed method was proved by coherence investigation of DC voltage group standard kept in the Central Military Calibration Laboratory (CWOM, Warsaw, Poland). Results obtained for the group of five Fluke DC voltage standards prove very good coherence of voltage differences results and confirmed accuracy of DC voltage unit in the CWOM at the level of 10-6.