Abstract

The paper presents the results of clinical trials using dynamic hip screw (DHS) for the treatment of fractures of the proximal femur and the results of numerical simulation of the model of femoral fixation using stabilizer DHS for the fixations most commonly occurring in clinical cases. The simulation results constituted the basis for the selection of mechanical properties and allowed to define the zones most vulnerable to mechanical failure of both the implant and bone tissue. Furthermore, designated zones of maximum stress enable the identification and assessment of corrosion processes in the areas of the damaged surface layer of the metal biomaterial.

Original languageEnglish
Title of host publicationAdvances in Intelligent Systems and Computing
PublisherSpringer Verlag
Pages124-131
Number of pages8
DOIs
Publication statusPublished - 1 Oct 2017

Publication series

NameAdvances in Intelligent Systems and Computing
Volume526
ISSN (Print)2194-5357

Keywords

  • DHS
  • Dynamic hip screw
  • Finite element analysis
  • Osteoporosis
  • Treatment of intertrochanteric fracture