Accounting for Overcounting Induced by 222Rn Contamination in 14C Measurements Performed with Liquid Scintillation Counting

Research output: Contribution to journal › Article › peer-review

1 Citation (Scopus)

Abstract

This research paper introduces a methodology for estimating overcounting resulting from uranium series radionuclides, with a specific focus on 222Rn contamination. The method is demonstrated using the Quantulus 1220TM spectrometer. Exponential fitting is employed for the calculations, utilizing a custom algorithm developed in MATLAB® R2022b. The proposed method allows for the determination of the sample count rate correction, F14C and radiocarbon age without the need for additional chemical treatment or specialized electronics during liquid scintillation counting (LSC) measurements with the Quantulus 1220TM spectrometer. However, the method does have certain limitations. It requires a significant number of cycles and extended measurement time per cycle, but it can be applied to samples with significant 222Rn contamination that require rapid 14C analysis.

Original languageEnglish
Pages (from-to)91-99
Number of pages9
JournalGeochronometria
Volume50
Issue number1
DOIs
Publication statusPublished - 1 Jan 2023

Keywords

  • Quantulus 1220spectrometer
  • Rn contamination
  • exponential fitting
  • liquid scintillation counting
  • overcounting estimation
  • uranium series radionuclides

ASJC Scopus subject areas

  • Earth and Planetary Sciences (miscellaneous)

Fingerprint

Dive into the research topics of 'Accounting for Overcounting Induced by 222Rn Contamination in 14C Measurements Performed with Liquid Scintillation Counting'. Together they form a unique fingerprint.

Cite this